Precautions for the use of inductively coupled plasma emission spectrometer

2021/2/25 16:55:20

Maintenance of gas control system by inductively coupled plasma emission spectrometer

         Whether the gas control system of ICP is operating stably and normally, directly affects the quality of the measurement data of the instrument. If there are water droplets in the gas path, mechanical debris and debris, etc., the gas flow will be unstable. Therefore, the gas control system must be frequently checked. Inspection and maintenance. First do a gas test. Turn on the power switch of the gas control system to make the solenoid valve work, then turn on the gas cylinder and the pressure reducing valve to make the gas pressure indicate the rated &#118alue, then close the gas cylinder and observe the pressure on the pressure reducing valve. The pointer of the pressure gauge should not drop or drop very little within a few hours, otherwise there will be air leakage in the gas path, which needs to be checked and eliminated. Second, because the argon gas is often mixed with moisture and other impurities, some mechanical debris in the pipes and joints will fall off, causing the gas path to be blocked. Therefore, it is necessary to clean up regularly, unplug certain sections of the pipeline, then open the gas cylinder, and release the gas for a short period of time to blow out the water droplets and dust particles in the pipeline. When installing the gas pipeline, especially when connecting the carrier gas pipeline to the atomizer, be careful not to bend the tube too much, otherwise the unstable carrier gas flow will cause pulsation and affect the measurement.

Maintenance of Inductively Coupled Plasma Emission Spectroscopy Instrument on the Sampling System and Torch

        The nebulizer is the most precise and critical part of the sampling system and requires good maintenance and use. Regular cleaning is required, especially after the determination of high-salt solutions, the top of the atomizer and the torch nozzle will accumulate salt, which will cause the aerosol passage to be unsmooth. This often reflects the decrease in the measurement intensity and the increase in the reflected power of the instrument. Dust or carbon deposits on the torch will affect the ignition of the plasma torch and its stability, and also affect the reflected power. Therefore, it should be washed regularly with acid, water, and finally, washed with absolute ethanol and blown dry. Always keep the sampling system and Cleaning of the torch.

Minimize the number of startup and shutdown during the use of inductively coupled plasma emission spectroscopy

        Before starting the measurement, you must make arrangements and mark all the preparations in advance. Do not start and stop during the same period of time. Frequent opening of the instrument can easily cause damage. This is because the instantaneous current is greatly high every time the instrument is turned on. The current during normal operation, the instantaneous pulse impact, will easily cause the filament of the power tube to break, short circuit and premature aging, etc. Therefore, you need to pay more attention to it in use. Once it is turned on, it will be done in one go. Shut down midway.

主站蜘蛛池模板: 黄平县| 阿荣旗| 九龙县| 巫山县| 富裕县| 井陉县| 兴仁县| 岱山县| 方城县| 林周县| 翁牛特旗| 甘泉县| 九龙城区| 浮山县| 三都| 衡山县| 晋中市| 弥勒县| 北海市| 枞阳县| 交城县| 措勤县| 新津县| 岢岚县| 岳阳县| 灵山县| 汉沽区| 龙南县| 中超| 烟台市| 新安县| 长白| 玉溪市| 北碚区| 班玛县| 洛南县| 梁山县| 宝清县| 蒙自县| 安徽省| 基隆市| 澜沧| 灵寿县| 五莲县| 隆昌县| 大悟县| 茌平县| 泰兴市| 遵义县| 九龙县| 咸丰县| 寻乌县| 苏州市| 中山市| 瑞昌市| 比如县| 泸州市| 三门县| 砀山县| 大安市| 永年县| 南召县| 陕西省| 岐山县| 阳山县| 周宁县| 谷城县| 武汉市| 汕尾市| 武汉市| 沙湾县| 灵丘县| 墨竹工卡县| 招远市| 巴中市| 延寿县| 揭东县| 文昌市| 临颍县| 三门县| 怀化市| 凤阳县|